Forum Post: RE: Respiration from body using ADS129xR EVM

$
0
0

Hi James, 

Thanks for checking out the ADS1298R!

Are you able to see the on-board respiration signal? This would require JP34 [1-2] and JP35 [1-2].

I found that the Respiration measurements with ADS129xR devices work best with ECG cables < 5 kOhms of impedance. Too much impedance will prevent the ADC from resolving the Respiration waveform.

Can you measure the impedance of your cables (from electrode to DB15 header)?

Best Regards,