Hi Thorsten, Thank you for all the information. I see what you are doing now. Keep in mind, as the analog input frequency changes from DC, the calibration may not hold up as well, due to BW mismatch between the two ADCs. If this is important to you and your application. What I would recommend, is using the ADC12DJ5200RF, it will work at this lower sampling rate and the gain, offset registers are exposed for further experimentation. Here is a link to the datasheet. www.ti.com/.../adc12dj5200rf.pdf Regards, Rob
↧