Forum Post: RE: ADC3664EVM: DCLK and Sampling Clock

$
0
0
Hi TTD, There is an error in the user guide. Please use a sinewave and set the level to 0 to +2dBm max. For the DCLKIN, you can use a function generator if this is clean enough. However, its best to use good quality low jitter (clean) signal generators. Some examples include the SMA100A or SMA100B from R&S, or Agilent 8257D. Any of these choices, needs to have the low jitter option installed. Each of these signal generators, needs to have a bandpass filter installed in-line from the output of the sig gen to the input SMA connector on the EVM. Here is a generic pic. Regards, Rob